Portable ECScope™ - 12i is the ECG Monitoring System from DyAnsys Inc. The equipment, with its accessories, is intended to acquire, analyze, display and record 12-lead electrocardiographic information.


Data acquisition

Standard leads acquired: I, II, III, aVR, aVL, aVF, V1 to V6
Lead and gain selection: User can dynamically select lead/gain at the start of recording
Digital sampling rate: 200 samples/sec
Sampling resolution: 12 bit, bipolar
Frequency response: 0.05 Hz to 150 Hz
   

Filters
High pass: 0.05Hz cutoff
Low pass: 150Hz cutoff
Notch filter: 50Hz, 60Hz (software filter)
Input dynamic range: ±5 mV
Common mode rejection: >100dB
DC offset capability: 300 mV DC
Input lead impedance: >20M ohm at DC
 
ECG Monitor
The ECG signals are displayed in real time on the LCD screen of the portable ECScope 12i.
The display also shows
      ECG lead selection
      ECG Gain selection
      Heart Rate
      ECG strips record option
      ECG interpretation
 
Patient safety
Patient leakage current: <10 uA
Classified as Class II - Type CF device as per IEC requirements
Protected from patient defibrillation
 
Physical specifications
Weight: <0.4Kgs (including the battery)
Dimensions: 75 x 95 x 112 (mm)
 
Environmental specifications
Operating temperature: 0°C to 45°C
Storage Temperature: –20°C to +65°C
Humidity: 10% to 95%
Pressure: 700 -1060 millibars
 
Regulatory compliance
IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC60601-2-25 and IEC 60601-2-27
 
Data storage capability
Dyansys USB memory key (Thumb drives) can be used for data storage/backup
Data can be exported to a PC via a USB memory key
 
Printing capability
Works with Dyansys specific printers
Printout includes patient demographics, 12 lead ECG graphs and interpreted data
Prints on A4/letter size papers in color providing an inexpensive printing solution
 
Power specifications
AC adapter input: 100-240V AC, 50/60Hz
AC adapter output: 12V DC @ 2amps Max
Power Consumption: <15W
 
Battery and charger
Battery pack type: NiMH rechargeable battery
Battery voltage: 3.6V DC
Battery capacity: 2500 mAh
Typical battery operational time on full charge (without any peripherals connected): >6hours
Typical battery charge time: 6 hours
 
Product Lifetime
5 Years
 
Patient Information
Input at the start of the measurement
 
Product Test and Calibration
Extended test and calibration can be performed including the ECG cable by connecting the ECG cables to calibration terminals provided in the product
 
Help Facility
Is available at any time on the display
 
Data Display
Displayed in color touch screen palmtop
 
Standard accessories
10 lead ECG cable
Product user manual
 
 


 
Note : specifications subject to change without notice.
News and Events
Dyansys Product Family
The Portable ANSiscope and ECScope Series for family physicians and field applications 
  more  
Dyansys Application
Diabetes
Anesthesiology
Cardiology
Measurement of autonomic dysfunction
  more  
Copyright © 2009 DyAnsys Inc. All Rights Reserved
Site designed by Uditi Technologies