Portable ECScope is a hand-held battery operated device to simultaneously monitor 7 lead ECGs. 6 repetitive measurements with chest lead at V1 to V6 provides a 12 lead ECG printout.
It can also simultaneously monitor 12 lead ECGs using 5 to 12 lead algorithm.
It is ideal for operation in general physician's office or in mobile telemetry.
Data acquisition
| No. of leads acquired: |
7 Leads simultaneous (Conventional mode) |
| |
12 Leads simultaneous (5 -12 mode) |
| Leads acquired (Conventional mode): |
I, II, III, aVR, aVL, aVF and V |
Leads acquired
(5 -12 mode): |
I, II, III, aVR, aVL, aVF, V1,V2,V3,V4,V5 and V6 |
| Lead and gain selection: |
User can dynamically select lead/gain during recording
|
| Digital sampling rate: |
200 samples/sec (5ms samples) |
| |
|
| Patient information
|
| Input at the start of the measurement |
| |
| Help facility
|
| Is available at any time on the display |
| |
| Data Display |
| Displayed in color touch screen |
| |
| ECG Monitoring and Recording |
| The ECG signals are displayed in real time on the LCD screen of the Portable ECScope. |
|
| The display also shows
|
| |
|
ECG lead selection |
| |
|
ECG gain selection |
| |
|
Heart Rate
|
| |
|
ECG strips record option |
|
| |
| Data storage capability |
|
Permanent storage of 200 patient records |
|
More than 2000 patient records can be stored in Dyansys |
| |
USB memory key |
|
Data can be transfered to a PC via the USB memory key |
|
| |
| Printing capability |
|
Direct Print out through the attached printer on a single A4 |
| |
sheet in plain or graph paper |
|
Print out using PC through Dyansys Print Tool |
|
Support 10 seconds Rhythm Print out |
|
Print out included Patient demographics, Doctors details, |
| |
12 Lead ECG graphs in format - 6 x 2 1R, 4 x 3 1R |
|
Email option using Dyansys Print Tool |
| |
|
|
| |
| Regulatory compliance |
| IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4 and IEC 60601-2-27 |
FDA Approved * |
| |
| Patient safety |
|
Patient leakage current: <10 uA |
|
Classified as Class II - Type CF device as per IEC requirements |
|
Protected from patient defibrillation |
|
| |
| Product test and calibration |
|
Extended tests and calibration can be performed including the
|
| |
ECG cable by connecting the ECG cables to calibration terminals provided in the product |
|
| |
| Power specifications |
|
AC adapter input |
230 V |
|
AC adapter output |
12V DC @ 2amps Max |
|
Power Consumption |
<15W |
|
| |
| Physical specifications |
|
Weight
|
<0.4 Kgs (including the battery) |
|
A
Dimensions |
175 x 95 x 112 (mm) |
|
| |
| Environmental specifications |
|
Operating temperature |
0°C to 45°C |
|
Storage Temperature |
–20°C to +65°C
|
|
Humidity |
10% to 95% |
|
Pressure |
700 -1060 millibars |
|
| |
| Battery and charger |
|
Battery pack type |
NiMH rechargeable battery |
|
Battery voltage |
3.6 VDC |
|
Battery capacity |
2500 mAh |
|
Typical battery operational |
>6 hours |
| |
time on full charge: (without any peripherals connected) |
|
|
Typical battery charge time |
6 hours |
|
| |
| Printer supported |
| HP 5168 |
| |
| ECG Electrodes |
| Use DyAnsys Electrodes for high signal quality |
| |
| ECG Cable |
| 5 lead ECG cable with locking connector |
| |
| |
|
| |
| Note
: specifications subject to change without notice. |
| |
|
|